Web: | mayocliniclabs.com |
---|---|
Email: | mcl@mayo.edu |
Telephone: | 800-533-1710 |
International: | +1 855-379-3115 |
Values are valid only on day of printing. |
Aldosterone-d6 is added to serum and plasma samples as an internal standard. Aldosterone and aldosterone-d6 are extracted from the specimens using a Strata X cartridge. The eluate is dried down under nitrogen, reconstituted with 70/30 methanol/H2O containing estriol and analyzed by liquid chromatography-tandem mass spectrometry using multiple reaction monitoring in the negative mode.(Fredline VF, Taylor PJ, Dodds HM, Johnson AG: A reference method for the analysis of aldosterone in blood by high-performance liquid chromatography-atmospheric pressure chemical ionization-tandem mass spectrometry. Anal Biochem 1997 Oct 15;252[2]:308-313)
Monday through Friday